Setup Phototroph eVOLVER

hashtag
Base eVOLVER Requirements

  1. From Fynch Bio:

    1. eVOLVER base

    2. Fluidics box

    3. 2X fast pump racks

    4. [Optional] Additional slow pump rack for inducers

  2. Dedicated computer for experimentation

    1. Mac strongly recommended for ease of setup and reliability

hashtag
Phototroph eVOLVER Hardware Setup

The following are different from the base eVOLVER :

  1. Additional fluidics

hashtag
OD Considerations

We pause stirring and phototroph growth lights when we take OD. Without this, OD would be nonfunctional. Stirring causes bubbles be pulled downwards and the growth lights completely saturate the OD sensor.

hashtag
Additional Calibration Procedures

  1. to calibration conf.yml before doing calibration

    1. Because this will give us data every 20 seconds and greatly speed up calibration

  2. Settings during calibration should mimic the experiment therefore:

Bubbling 1-16 millifluidic splitter
  • Gas mixer (or tank of premixed gas with regulators)

  • Vials (recommended to get parts for at least two sets)

    1. 5-port nylon tube cap

    2. Bubbler (plugs in to cap)

    3. Long thin stir bars - required otherwise bubbles will stick and throw off OD readings

  • Bubbler cleaning apparatus (recommended for saving time)

  • Light meter and probe

    1. Our light meter is the LI-COR LI-1500arrow-up-right

    2. Our light probe is the Walz US-SQS/L omnidirectional light probearrow-up-right

  • Phototroph growth lights off

  • Stirring off

  • Vial setup for calibration (also mimic the experiment)

    1. Prepare standards in caps with a bubbler

    2. Use long stir bars

  • When moving standards, give a swirl once every other time

    1. This is to make sure cells don’t settle and change the OD

  • Finally, switch back to experiment conf.yml when ready for experiment

  • Router
    setup
    Smart sleeve construction
    Emergency efflux
    Switch