Optical Density (OD) Readings

hashtag
Overview

Optical density (OD) is one of the most integral parts of continuous culture. While chemostats do not innately rely on OD, it is still a useful metric for your evolution. Meanwhile, turbidostats and growth curves heavily rely on OD to function.

However, OD is the most easily disrupted parts of eVOLVER hardware. This guide serves as an initial list of things to check when troubleshooting OD.

circle-exclamation

Disclaimer: You may find that you are still not getting "good" OD readings after trying solutions in this guide. We hope that you will search the because others may have had your same issue.

circle-exclamation

After trying a fix, before attempting a whole new calibration, prepare a couple of standards in the OD range that you care about and check if you see a difference between them in the raw OD values (in the GUI Setup page or by starting an experiment and going to the data folder).

circle-info

For more in-depth information about the OD hardware see .

hashtag
Most Common Problems

hashtag
Variability in OD Sensing Range Between Vials

  1. If you already have a version of the 3D printed tube holder with defined OD LED and photodiode distances:

    1. Use the upgrade to double check your photodiodes and LEDs are correctly placed!

hashtag
Noisy OD Data

  1. Stir bars jumping around during OD readings

    1. Decrease stir speed

hashtag
Incorrect OD

ie eVOLVER OD is offset from the cuvette or plate reader measured value.

  1. Bad

    1. Especially because you did not wait for temperatures of the vials to equilibriate

  2. Poor quality OD calibration

hashtag
No Change in OD

  1. Photodiode or IR LED not working

    1. Replacement

    2. A failed photodiode will show a raw OD value close to 64,000

hashtag
Other Considerations

Pre-Calibration Considerations

  • od_led value in conf.yml file (PWM value)

Experiment Considerations

  • Photodiode Temperature - make sure temperature is equilibrated before calibration and before start of experiment

  • eVOLVER recently turned on - causes rapid change in values

  • -

Non-standard changes

  • Vial holder reflectiveness - ie white vs black vs metallic

Sometimes a given OD photodiode or LED can be variable as they are cheap parts. Swap both out for new ones. Part numbers .
  • Try

  • Try different ADC resistor packs

    1. eVOLVERs come with a variety of these

    2. Ask on the forum for help

  • Check that your vial construction is not variable

  • You are near the edge of your calibrated OD range
  • is off enough to make your calibration fail

  • You can check your OD calibration using the script
  • Different temperatures require different OD calibrations

  • A failed IR LED will most likely show a raw OD value much closer to your other vials
  • Check that they have the long (positive) lead in the correct positive terminal

  • ADC board or motherboard broken

    1. Can happen after vial overflow

    2. See motherboard and

    • Use the splash guard that comes with your eVOLVER

    • Or using black acrylic

    • You may need to recalibrate if you calibrated with ambient light

    Needles or other structures coming below 10mL line (close to OD PD etc)

  • Vial volume - a volume of 10mL will totally change OD compared to 30

  • Stirring with low vial volume - this will give noisy readings as the funnel created by stirring fluctuates and is more and less reflective

  • forum arrow-up-right
    here
    Standardize LED/Photodiode distance
    guide
    Stop stirring during OD readingsarrow-up-right
    OD blank
    guide
    LED to photodiode anglearrow-up-right
    Resistor pack valuearrow-up-right
    Vial to Vial Sleeve Distancearrow-up-right
    Second Forum Postarrow-up-right
    Low (~10mL) culture volumes ONLY - readings are affected by stirringarrow-up-right
    Vial 1 (left) shows a smaller range than vial 2 (right). However, vial 2 is still unable to accurately differentiate values under OD 0.5.
    here
    OD90 for OD600 0.5-4.5 or OD135 for OD600 0.0-0.5
    OD blank
    manual calibration
    hardware
    troubleshooting
    Ambient lightarrow-up-right
    make onearrow-up-right