arrow-left

All pages
gitbookPowered by GitBook
1 of 9

Loading...

Loading...

Loading...

Loading...

Loading...

Loading...

Loading...

Loading...

Loading...

Using the extractor script

hashtag
Step-wise Process for Setting Up the Script for Extractor Experiments

1. Select the Type of Experiment

  • Decide whether to run a Chemostat or Turbidostat experiment.

  • Ensure you choose the correct type when configuring the script.

2. Set Parameters for the Experiment

  • The setup process for Chemostat and Turbidostat is identical to non-extractor versions.

  • Define the operational parameters as you normally would for the chosen type of experiment.

3. Specify Extractor and Non-Extractor Vials

  • Identify which vials will be used as extractor and non-extractor vials:

    • Common Extractor Vials: 4, 5, 6, 7, 12, 13, 14, and 15.

    • Common Non-Extractor Vials: 0, 1, 2, 3, 8, 9, 10, and 11.

4. Edit Vial Assignments if Needed

  • If the vial assignments need to be changed during the experiment:

    • Modify the script accordingly.

    • Ensure all changes are accurately reflected in the script.

5. Restart the Script After Changes

  • If any changes are made to the extractor or non-extractor vial assignments:

    • Stop the current running script.

    • Restart the updated script in the terminal to apply the changes.

6. Run the Experiment

  • Once the script is set up with the correct parameters and vial assignments:

    • Start running the experiment using the script.

7. Monitor and Adjust as Necessary

  • During the experiment, monitor the setup to ensure everything functions as expected.

  • If adjustments are needed, repeat the steps to edit the script and restart it.

Edit the script to assign these roles to the corresponding vials.

Extractor setting in script

Troubleshooting

Possible errors you can run into

The main result of issues with the extractor can lead to extractor or non-extractor vials to being empty or result in overflow.

Pair of vials where extractor vial was all pumped out

Some factors that can contribute to this are:

Condensation:

Condensation can occur in the extractor vials depending on ambient conditions in the extractor as well as what organism you are working with. Condensation can obscure the photodiodes tricking them into thinking the vial has more volume then it actually does, triggering unwanted pump events. Generally the best solution for this is to ensure the heat is properly working or to raise the temperature of the vial.

Disturbing the experiment:

The extractor and non-extractor vials are delicate when the experiment is actively running. Vials should NOT be moved or sampled while the experiment is actively running. It is recommended to pause the experiment before handling any vials. Moving the extractor vials while the experiment is running can negatively impact the fluidics set up, or move the liquid into range of the photodiodes triggering an unwanted pump event.

Contamination:

Contamination of vials:

Vial contamination will cause a rapid increase in OD which will rapidly accelerate the rate at which pump events occur.

Contamination of media bottle:

If the non-extractor vial begins to overfill if it reaches the main media influx line the inoculated media can work its way back through the line towards your media bottle. Additionally the same can occur if you have your vial-to-vial line dispensing liquid from above your main media influx line which will again expose the main media line to inoculated media.

Improper of the fluidic lines can also cause contamination

Incorrect Fluidics setup:

Incorrect set up of the fluidics can cause immediate problems during your experiment make sure to double check your before starting your experiment

Droplets on side of vial:

In the extractor vials if the straw dripping culture in is placed too high above the liquid level the drops from the straw can splash and cause droplets to collect on the side of the vials. Causing a similar effect to condensation by obscuring the photodiodes triggering unwanted pump events. When designing your extractor caps measure the straw length so it sits at about 1 inch above your desired volume.

Incorrect volume/block setup:

Remember to double check what volume you are putting into the extractor vial, currently the volume of the extractor vials should be no greater than 15 mL the photodiodes should blank "nothing" at both blanking steps extractor vials exceeding 15 mL can interfere with the blanking steps causing immediate problems with pump events. Similarly if using acrylic blocks to elevate extractor vials remember to not account for how much "volume" each block take up.

Dirty vials:

Improperly cleaned vials can run into the same issues as condensation or droplets on the vial, again tricking the photodiodes into triggering an unwanted pump event.

sterilization
setup

Extractor Analysis

Sterilizing Extractor Fluidics

How to set up the extractor fluidics utilizing the slow pump systems

  • Set up your lines using the as the traditional eVOLVER

  • Use special method to run the slow pumps and fill them for sterilization

same method
this

Extractor Column

hashtag
Overview

The two column extractor is a type of eVOLVER experiment designed to continuously extract a molecule of interest exuded by a microbe. One set of vials are bioreactors where cells are continuously cultured. Those vials are each paired with an extractor vial, where microbes are continuous dropped through a solvent layer that extracts a biomolecule of interest. This solvent can be sampled to determine amount of biomolecule that is made.

Diagram of extractor fluidics.

Experiment Start

Setting up and starting an experiment with the extractor set up

circle-exclamation

Calibrate your eVOLVER before beginning your experiment

When starting an experiment you'll go through the same process as a normal experiment. Make sure that you have the extractor columns properly set up and that you also include the slow pumps lines in your line sterilization process.

circle-info

Preemptively check the troubleshooting to see common mistakes before running the experiment

page

Example protocols

Sample protocols of experiments ran using the two column extractor

Two-week Extractor experiment

Turbidostat extractor experiment

Chemostat extractor experiment

Feedstock experiment

https://docs.google.com/document/d/1akdpiJPpmqBr1jAJhz33LwnbTb1YgxgTWrXEN6bRQqU/edit?tab=t.0docs.google.comchevron-right

Extractor Volume Maintenance

hashtag
Maintaining the Solvent Layer

The solvent layer on top of the cells / media in the extractor must be maintained. However, in the basic eVOLVER setup, volume in the vial is maintained by setting the efflux straw's height to a certain vial volume and pumping any excess volume away. Therefore in basic eVOLVER, the solvent layer would be the first to go.

Instead, the extractor controls the volume in the vial by using the od90 photodiode - LED pair to monitor fluid levels.

hashtag
Volume Maintenance Algorithm

  1. Experiment started

  2. Pump Wait: No pumping until vol_check_wait is exceeded

  3. Blanking: the value where the volume line is below the volume sensor is taken

The raw od90 values just before vol_check_wait are stored and averaged to use as this threshold 'blank'

  • Continuous Pumping: Cells are pumped from the cell growth bioreactor (chemostat or turbidostat) to the extractor and back from the extractor to the cell growth bioreactor

    1. In cycles based on the eVOLVER broadcast timing

  • Pump Duration Biasing: Pumping out from the extractor is shorter than pumping in

  • Extractor Volume Increases: The volume in the extractor steadily rises towards the volume sensor

  • Volume Threshold Exceeded: When the volume line increases above threshold, the system will only pump OUT one broadcast cycle, decreasing the volume level

  • Steps repeat

  • vol_check_wait = 0.15 # (hours) time to wait before regulating extractor volume

    Setting up your experiment

    1. Prepare and Organize Your Equipment

    • Gather all necessary vials, slow pumps, tubing, and any other required equipment.

      • Recommended equipment: clean vials, pumps, tubing, 1:1 extractor/non-extractor caps, appropriate media, clean beakers, 10% Bleach, 70% Ethanol

    • Ensure that each extractor vial has a matching non-extractor vial (e.g., vial pair 0 and 4).

    2. Connect the Tubing

    • Match each pump with one extractor/non-extractor pair.

    • For each pair, connect the tubing from the pump to the appropriate vials:

      • For example, if using vial pair 0 and 4, connect the pump to suck up liquid from vial 0 and dispense it into vial 4.

    3. Check Tubing Pathways

    • Carefully follow the path of the tubing:

      • Ensure the tube is correctly drawing from the extractor vial (e.g., vial 0) and dispensing into the non-extractor vial (e.g., vial 4).

      • Similarly, ensure the tube for the reverse flow (non-extractor to extractor) is correctly set up.

    While the tubing set up may look confusing at a glance remember that logically the setup works the same and as long as you understand the job of each pump on the slow pump array and the port to each straw on the cap you can trace back what goes where

    triangle-exclamation

    When setting up the media influx as well as the vial-to-vial efflux in your Chemostat/Turbidostat vial ensure the media influx line is positioned ABOVE the vial-to-vial influx failure can result in contamination of media bottle.

    The two angled straws positioned closer to the cap are your media influx and vial-to-vial efflux

    4. Volume and Media Preparation

    • Prepare 20 mL of media for the Chemostat/Turbidostat vials used in the experiment.

    • For extractor columns, use a media volume of 15 mL to prevent the liquid level from reaching the photodiodes/LEDs.

    • Use the pumps/slow pumps to fill lines with media

    5. Positioning of Vials

    • Ensure that the vials are placed correctly making sure you account for any added acrylic blocks underneath.

    • Double-check that the liquid levels are BELOW the photodiodes/LEDs.

    6. Monitor the Pumping Process

    • Observe the pumping operation closely:

      • Ensure that the correct volumes are being transferred between vials as intended.

      • Monitor for any signs of malfunction or unexpected behavior.

    triangle-exclamation

    Make sure to observe the straws carefully to ensure they are filled with media. Below is an example of a vial where some straws still have pockets of air.

    7. Ensure Cleanliness of Vials

    • Clean vials thoroughly to avoid stains, condensation, or drippage.

    • Make sure there are no residual liquids that can impact the sensor readings.

    9. Adjust if Necessary

    • Make adjustments to tubing, pumps, or vial positions if liquid levels are incorrect or if the system does not operate as expected.

    9. Start the Experiment

    • Once all the checks and setups are completed, start the experiment with your .

    • Regularly monitor the system to ensure it is functioning correctly throughout the experiment.

    circle-info

    The extractor will perform 2 blanks. One directly after starting the experiment and another 15 minutes into the experiment, the second blank can be changed in the .

    By following these steps, you can ensure a successful setup and operation of the two-column extractor system, minimizing errors and ensuring accurate experimental results.

    Fill non-extractor lines and vials with media first and using a secondary set of extractor vials with media in them already second to prevent bubbles from forming in slow pump lines
    • This is important because the initial blank should be based off of filled straws

    extractor script
    extractor script
    Vial Top View
    Vial Side View
    Diagram of extractor fluidics setup.
    Close up view between Vial 0 (non-extractor) and Vial 4 (extractor)
    Front View of just vials 0 (Left) and 4 (Right) outside of smart sleeve
    Side view of just vials 0 and 4 on eVOLVER
    Secondary Extractor Vial (Left) Experimental Extractor Vial (Middle) Non-Extractor Vial (Right)
    Extractor Vial with straws properly filled (Ready for blank)
    Acrylic blocks underneath can change the positional height of the vial
    run the slow pumps