Bubbler Cleaning / Testing Apparatus
Overview
This apparatus is designed to save you time when you are cleaning bubblers.

Construction
Parts
Infrastructure (You may already have this in place)
- 1X - Gas regulator with pressure gauge capable of regulating to 10 psi - For controlling air pressure to bubblers (we don't want 150 psi going through our bubblers!) 
- 1X - Push-to-Connect Tube 1/8" Tube OD x 1/8 NPT Male - for going to nylon 1/8" tubing 
 
- 1X - Vacuum trap - For example something like a filter flask before the vacuum line 
 
Make sure to have a vacuum trap to prevent fluids from going into your vacuum line! You don't want facilities mad at you!
Cleaning Apparatus
- Teflon tape - sealant for leaky threaded fittings (tightly wrap it around and break it off) 
- At least 6ft - Nylon tubing (1/8" OD) 
- Flexible silicone rubber tubing (3/32" ID (inner diameter)) - Alternatively you can try and jam some extra 1/16" efflux tubing onto the 1/8" nylon tubing 
 
- 50mL falcon tube cap (or an extra eVOLVER cap; see below for usage in push-to-connect cap) 
Assembly

1. Assemble apparatus
- Assemble the 3 push-button connectors using directions that come with them 
- Connect nylon tubing in the above configuration 
- Connect 3/32" tubing in the above configuration 
2. Assemble a vacuum trap to prevent liquids from entering building vacuum line
3. [Bubbler testing only] Assemble the push-to-connect cap
Used for easily attaching / detaching bubblers for testing

Testing Protocol
Never allow water, bleach, or media into the apparatus as it will corrode the valves in the push-button connectors.
- Attach the bubbler to the black 2-way connector connected to the green falcon tube lid as pictured below. To attach, push the bubbler in; to detach, push the black ledge into the connector and then pull the bubbler out. 

- Fill a vial with ~15mL of rich media like LB, YPD, or BHI. BG11 media will also work. As a rule of thumb, the less salt and peptides your media has dissolved in it, the larger your bubbles will be. 
- Place the green lid with the bubbler attached into the vial with media. Ensure the bubbler is submerged in the liquid. 

- Ensure the air supply is turned on and set to 5-10psi. Check that the bubbler cleaning apparatus tubing is attached to the air supply. If air is flowing through the apparatus, you should feel a rush of air coming out of the red tubing. 