Bubbler Cleaning / Testing Apparatus
Overview
This apparatus is designed to save you time when you are cleaning bubblers.
Construction
Parts
Infrastructure (You may already have this in place)
1X - Gas regulator with pressure gauge capable of regulating to 10 psi
For controlling air pressure to bubblers (we don't want 150 psi going through our bubblers!)
1X - Push-to-Connect Tube 1/8" Tube OD x 1/8 NPT Male - for going to nylon 1/8" tubing
1X - Vacuum trap
For example something like a filter flask before the vacuum line
Make sure to have a vacuum trap to prevent fluids from going into your vacuum line! You don't want facilities mad at you!
Cleaning Apparatus
Teflon tape - sealant for leaky threaded fittings (tightly wrap it around and break it off)
At least 6ft - Nylon tubing (1/8" OD)
Flexible silicone rubber tubing (3/32" ID (inner diameter))
Alternatively you can try and jam some extra 1/16" efflux tubing onto the 1/8" nylon tubing
50mL falcon tube cap (or an extra eVOLVER cap; see below for usage in push-to-connect cap)
Assembly
1. Assemble apparatus
Assemble the 3 push-button connectors using directions that come with them
Connect nylon tubing in the above configuration
Connect 3/32" tubing in the above configuration
2. Assemble a vacuum trap to prevent liquids from entering building vacuum line
3. [Bubbler testing only] Assemble the push-to-connect cap
Used for easily attaching / detaching bubblers for testing

Testing Protocol
Never allow water, bleach, or media into the apparatus as it will corrode the valves in the push-button connectors.
Attach the bubbler to the black 2-way connector connected to the green falcon tube lid as pictured below. To attach, push the bubbler in; to detach, push the black ledge into the connector and then pull the bubbler out.

Fill a vial with ~15mL of rich media like LB, YPD, or BHI. BG11 media will also work. As a rule of thumb, the less salt and peptides your media has dissolved in it, the larger your bubbles will be.
Place the green lid with the bubbler attached into the vial with media. Ensure the bubbler is submerged in the liquid.

Ensure the air supply is turned on and set to 5-10psi. Check that the bubbler cleaning apparatus tubing is attached to the air supply. If air is flowing through the apparatus, you should feel a rush of air coming out of the red tubing.