Manual Calibration - calibrate.py

circle-info

You can use manual calibration to view past calibrations after they are done. This can be useful for troubleshooting if a vial is not behaving as expected.

About

The GUI will run calibrations automatically upon completion of the calibration protocols. However, you can still manually run a calibration if you would like to change calibration settings. Can be useful if your GUI calibration fails for some reason.

Located in /dpu/calibration/arrow-up-right.

Requirements

  1. You have run calibration and logged raw values already

  2. You are in your DPU virtual environment in command line. This is set up when you install the DPU.

Commands

List raw calibration files on the eVOLVER

Mac

python3 calibration/calibrate.py -a <ip_address> -g

For Windows, use py instead of python for all commands.

Calibrate Temperature

Calibrate OD

OD135

OD90 (Check to ensure mode is configured properly)

3D FIT (Check to ensure mode is configured properly)