eVOLVER
  • eVOLVER Documentation Wiki
  • General
    • About Us
    • eVOLVER Community
      • Code of Conduct
  • Getting Started
    • Buying eVOLVER
    • Part Sourcing
    • Unboxing and Setting Up
    • Software Installation
      • DPU Installation
      • Electron App (GUI) Installation
    • Configuring Computer and Networking
      • Router Setup
    • Calibrations
      • Temperature Calibration
      • Optical Density Calibration
      • Pump Calibration
      • Manual Calibration - calibrate.py
  • Upgrade Base eVOLVER Hardware
  • Experiments
    • Starting an Experiment
      • Carboy Media Prep
      • Preparing Vials
      • Setup Waste Carboy
      • Sterilizing Lines
      • Loading Vials and Setting Initial Conditions
      • OD Blank
      • GUI Start Guide
      • Command Line Start Guide
      • Cleaning Up After Experiment
    • Growth Curve
    • Chemostat
    • Turbidostat
    • FAQs
    • Tips and Tricks
  • Guides
    • Use the GUI to Control Parameters
    • Building a Smart Sleeve
    • Making media bottles and splitters
    • Emergency Efflux
    • View the Server Log and Restart Server
    • Updating the eVOLVER Server
    • Change Your conf.yml File
    • Arduino Software Installation
    • Raspberry Pi Configuration
    • Command Line Usage
    • Millifluidics Guides
      • Designing Millifluidics Using Eagle
      • Constructing Laser Cut Millifluidics
      • Constructing Millifluidics via SLA Printing
      • Calibrating IPPs
      • Operating Millifluidics
  • Troubleshooting
    • Troubleshooting Overview
    • Experiment Troubleshooting
      • Contamination
      • Vial Overflow, Pump Failure, and Spills
      • Tubing and Connector Blockage and Bursting
    • Optical Density (OD) Readings
    • Vial Troubleshooting
      • Replacing Photodiodes and LEDs
      • Heating Element
      • Stirring
    • Server Troubleshooting
    • Vial Platform Troubleshooting
      • Motherboard Troubleshooting/Replacement
      • 12V Power Supply Troubleshooting/Replacement
    • GUI Troubleshooting
    • eVOLVER Maintenance
  • Hardware
    • Overview of Hardware Architecture
    • Overview of Fluidics
      • Tubing and connectors
      • Peristaltic Pumps
      • Fluidics box
    • Overview of Millifluidics
      • IPPs (Integrated Peristaltic Pumps)
      • Pressure Regulator
    • Vial Caps
      • Universal Vial Cap Construction Guide (Luer Connectors Only)
      • 5 and 7 Port Nylon Tubing Caps Construction Protocol
    • Smart Sleeve
      • Vial Board
      • 🌪️Stirring
      • Temperature
      • Optical Density
        • OD90 vs OD135
    • Motherboard Layout and Circuitry
      • 🌡️Arduino
      • Sensor/Actuator Board Slots
      • Pulse Width Modulation (PWM) Boards
      • Analog-to-Digital Converter (ADC) Boards
      • RS485 Board
    • Raspberry Pi
    • Chassis
    • Light Blocker / Splash Guard
    • Known Issues
  • Software
    • Overview of Software Architecture
    • DPU
      • Calibration
      • custom_script.py
      • Experiment Data Files
      • eVOLVER.py
    • Arduino
    • Server (Raspberry Pi)
      • Calibration Files
      • Configuration Files (conf.yml)
    • Known Issues
  • Extensions
    • Adding A New Experimental Parameter
      • Power Supply
      • Specific Applications
      • Custom Calibration Code
    • Custom Experiments
      • ePACE
        • ATTiny1634 Writing
        • LUX Board Troubleshooting
      • Morbidostat
      • Extractor Column
        • Extractor Volume Maintenance
        • Experiment Start
          • Sterilizing Extractor Fluidics
          • Setting up your experiment
          • Using the extractor script
        • Extractor Analysis
        • Troubleshooting
        • Example protocols
      • Phototroph Growth
        • Setup Phototroph eVOLVER
        • Photo-eVOLVER Smart Sleeves
          • Photo-eVOLVER Smart Sleeve Construction Guide
        • Experiment Guide
    • Custom Fluidics
      • Adding a Third Pump Rack
      • Bubblers / In-Vial Aeration
        • Bubbler Construction Protocol
        • Bubbler Cleaning Protocol
      • Running the slow pumps
    • min-eVOLVER
      • About
      • min-eVOLVER Construction
        • Parts
        • Construction Protocol
      • Fluidics Setup
      • Software Installation and Startup
      • send_command.py
      • Calibrations
      • Starting an Experiment
      • ePACE with min-eVOLVER
        • [v1.1] ePACE with min-eVOLVER
      • Troubleshooting
    • Interfacing with Other Systems
  • Contributing
    • Guidelines for Contribution
    • Reporting a Bug / Hardware Failure
    • Documentation
      • Making a Forum Post
      • How to Edit the Wiki
    • Software Development
    • Hardware Development
Powered by GitBook
On this page
  • About
  • Materials
  • Methods

Was this helpful?

Edit on GitHub
Export as PDF
  1. Getting Started
  2. Calibrations

Temperature Calibration

The goal of this tutorial is to learn how to calibrate the temperature settings for each Smart Sleeve in a 16 vial eVOLVER setup

PreviousCalibrationsNextOptical Density Calibration

Last updated 3 months ago

Was this helpful?

You must complete temperature calibrations before doing OD calibrations. Temperature affects OD and if there are big differences between vials, you will need to recalibrate all vials once you do a temperature calibration.

About

Why are temperature calibrations important?

Each microorganism has its own culture temperature requirement. Temperature may affect the successful cultivation and growth characteristics of the culture. By using a simple thermistor and heating resistor pair, we can track and change the culture temperature over time.

Why is calibration necessary?

Readings sent from the Raspberry Pi to the computer/server are the raw thermistor values measured by the Arduino. To make sense of the readings in terms of values we understand, we need a calibration file that converts this to degrees (celsius).

Do I need to redo temperature calibration if I switch out a thermistor/resistor pair?

Yes. Variability exists across the individual thermistor and heating resistor elements, necessitating a new temperature calibration anytime one of these parts are replaced. This is why the calibration curve is unique to each smart sleeve.

Materials

  • 16 eVOLVER vials filled with 15mL water

    • Glass vials - 40mL ()

  • Thermometer with probes

    • Do not use a thermocouple-based thermometer, as the magnetic field from the stir components can interfere with measurements

    • We use the ""

  • eVOLVER Electron GUI

    • This tutorial is for the newest version of the eVOLVER Electron App (Release 2.0.1), which enables both data visualization between each temperature calibration step (RT -> +10C -> +20C), as well as GUI-enabled calibration (without use of the command line).

    • For older versions of the Electron GUI (Release v1.2.1 and earlier), additional notes are provided at the bottom to successfully complete the calibration - the process remains largely the same, apart from the use of the command line to complete the calibration.

  • eVOLVER

Methods

Temperature Equilibrium

Generally, there are two independent methods to determine if the vial water has reached temperature equilibrium.

  1. Place the thermometer probes into the water, clear the "Min/Max" readings, and wait until the reads no longer fluctuates within ~2min of monitoring.

  2. The eVOLVER Electron GUI has an in-built function that displays the change in thermistor value from the last read, shown as a (+/-) value in brackets next to the current thermistor value being read. The GUI uses this information to determine whether the vials have equilibrated across time, shown as an orange circle indicator around each vial circle on the GUI that should turn fully orange upon equilibration.

    • Importantly, sometimes the vials will not display the achievement of equilibrium correctly. Please use your discretion with respect to the time elapsed under heat, lack of temperature fluctuations, similarity of temperature measured by thermometer across vials, and the GUI equilibrium function to determine when equilibrium has been reached.

  • Turn on the eVOLVER 5V power supply. Wait 5 seconds. Turn on the eVOLVER 12V power supply.

  • Open up the eVOLVER Electron GUI app, and ensure you are connected to the correct eVOLVER you wish to calibrate in the upper-right hand corner of the "Home Screen". The circle should be solid green, indicating that the eVOLVER is connected.

  • Tap "Setup" in the GUI home screen, select all vials by tapping "Select All" in the bottom right corner. The vial circles should highlight orange. Tap the "Next ->" button in the upper right-hand box (which includes all the control parameters) until you reach the "Temperature" tab. Drag the slider all the way down to "20C". Tap the button, and the "Executed Commands" box in the lower right should indicate that a temp command was sent to all vials.

  • Place all the eVOLVER vials, pre-filled with 15mL of water, into the Smart Sleeves.

  • Go back to the "Home Screen", click "Calibrations" and then "Temp" in the eVOLVER Electron App.

  • Name the calibration, hit enter, and then click "Start Temperature Calibration". A prompt will read "Heaters turned off. Let equilibrate, then enter values."

  • Room Temperature: Wait ~30 minutes for the temperature of the water to equilibrate to ambient room temperature (and heat generated by the eVOLVER).

  • Once equilibrated, measure the temperature of each vial with the thermometer probe and log the temperature measurements in Celsius in the Electron GUI.

The newest version of the eVOLVER Electron App (Release 2.0.1) enables data visualization between each temperature calibration step (RT -> +10C -> +20C). This is especially useful when verifying that the thermistor reads are as expected, and follows an inverse linear relationship between thermistor value and temperature.

  • Wait 60-120 minutes for the temperature of the water to equilibrate. Use the system outlined in the section "Temperature Equilibrium" to establish when the temperature has equilibrated.

  • Wait ~2-3hrs, or leave overnight.

  • Calibrate: An edit icon button should appear once all 3 measurements have been made. Click on the button, and a graph of the linear calibration curves for all thermistors across the smart sleeves should appear. Please confirm that the curves look linear. If everything looks good, please exit out of the graph (press the backwards arrow twice), and voila! You have completed your Temperature Calibration! Please be sure to change the temperature calibration file to the most recent in the eVOLVER set-up menu (click on the temperature calibration button).

IF In-GUI Calibration Fails (Unexpected Javascript/Python Error)

If after clicking on the "Edit" button upon completion of all 3 measurements, the eVOLVER GUI encounters an unexpected Javascript/Python error, do not fret! The raw calibration data is now saved on the Raspberry Pi, so you just need to continue along to the info note below to run the calibration through the command line. Feel free to exit the calibration at this point - your collected data is safe.

For older versions of the eVOLVER electron GUI (Release v1.2.1 and earlier)

If you are using an older version of the eVOLVER electron GUI (< v2.0.0), an additional step on the command line/terminal is required to linearly fit the acquired calibration data and set-up the temperature calibration. Additionally, there have been instances where unexpected code errors arise, necessitating calibration through command line. You can also still manually run a calibration if you would like to change calibration settings.

python calibration/calibrate.py -a <ip_address> -g

For Windows, use py instead of python for all commands.

Now that you have the name of the temperature calibration file, manually run the calibration file:

python calibration/calibrate.py -a <ip_address> -n <file_name> -t linear -f <name_after_fit> -p temp

A figure will pop up displaying the linear temperature calibration curves for each vial. If all looks good (inverse linear relationship between thermistor and temperature values - please see above figure for an example of a successful temperature calibration), then exit out of the figure. A prompt on the command line will ask if you would like to update the eVOLVER to the new temperature calibration. Press and enter "y", and voila, you are all done with temperature calibration.

Press the "RT " button and wait for the eVOLVER to measure the thermistor values (the eVOLVER measures the thermistor values continuously for a minute, with an orange ring appearing around the button. Once complete, the GUI should display "1/3 Measurements made"

Room Temperature +10C: Once complete, click the " +10C ->" button, and then click the "RT +10C ". The GUI should display "Temperature set. Let equilibrate, then enter values."

Once equilibrated, measure the temperature of each vial with the thermometer probe and log the temperature measurements in Celsius in the Electron GUI. Press the "RT +10C " button and wait for the eVOLVER to measure the thermistor values. Once complete, the GUI should display "2/3 measurements made"

Room Temperature +20C: Once complete, click the " +20C ->" button, and then click the "RT +20C ". The GUI should display "Temperature set. Let equilibrate, then enter values."

Once equilibrated, measure the temperature of each vial with the thermometer probe and log the temperature measurements in Celsius in the Electron GUI. Press the "RT +20C " button and wait for the eVOLVER to measure the thermistor values. Once complete, the GUI should display "3/3 measurements made"

Open up a terminal, and navigate to the eVOLVER dpu folder ( for download). List the raw calibration files on the eVOLVER Pi using:

▶️
▶️
▶️
▶️
▶️
▶️
Chemglass, CG-4902-08
Fisherbrand™ Traceable™ Excursion-Trac™ Datalogging Thermometers
github link
Real-time temperature calibration data visualization on the eVOLVER GUI interface
Example image of a successful temperature calibration