Manual Calibration - calibrate.py

Advanced Users Only:

Currently called by the GUI

Use GUI for calibrations in most cases

About

The GUI will run calibrations automatically upon completion of the calibration protocols. However, you can still manually run a calibration if you would like to change calibration settings. Can be useful if your GUI calibration fails for some reason.

Located in /dpu/calibration/.

Requirements

  1. You have run calibration and logged raw values already

  2. You are in your DPU virtual environment in command line. This is set up when you install the DPU.

Commands

List raw calibration files on eVOLVER

Mac

python3 calibration/calibrate.py -a <ip_address> -g

For Windows, use py instead of python for all commands.

Calibrate Temperature

python3 calibration/calibrate.py -a <ip_address> -n <file_name> -t linear -f <name_after_fit> -p temp

List raw OD JSON files logged on evolver

OD135

python3 calibration/calibrate.py -a <ip_address> -n <file_name> -t sigmoid -f <name_after_fit> -p od_135

OD90 (Check to ensure mode is configured properly)

python3 calibration/calibrate.py -a <ip_address> -n <file_name> -t sigmoid -f <name_after_fit> -p od_90

3D FIT (Check to ensure mode is configured properly)

python3 calibration/calibrate.py -a <ip_address> -n <file_name> -t 3d -f <name_after_fit> -p od_90,od_135