eVOLVER
  • eVOLVER Documentation Wiki
  • General
    • About Us
    • eVOLVER Community
      • Code of Conduct
  • Getting Started
    • Buying eVOLVER
    • Part Sourcing
    • Unboxing and Setting Up
    • Software Installation
      • DPU Installation
      • Electron App (GUI) Installation
    • Configuring Computer and Networking
      • Router Setup
    • Calibrations
      • Temperature Calibration
      • Optical Density Calibration
      • Pump Calibration
      • Manual Calibration - calibrate.py
  • Upgrade Base eVOLVER Hardware
  • Experiments
    • Starting an Experiment
      • Carboy Media Prep
      • Preparing Vials
      • Setup Waste Carboy
      • Sterilizing Lines
      • Loading Vials and Setting Initial Conditions
      • OD Blank
      • GUI Start Guide
      • Command Line Start Guide
      • Cleaning Up After Experiment
    • Growth Curve
    • Chemostat
    • Turbidostat
    • FAQs
    • Tips and Tricks
  • Guides
    • Use the GUI to Control Parameters
    • Building a Smart Sleeve
    • Making media bottles and splitters
    • Emergency Efflux
    • View the Server Log and Restart Server
    • Updating the eVOLVER Server
    • Change Your conf.yml File
    • Arduino Software Installation
    • Raspberry Pi Configuration
    • Command Line Usage
    • Millifluidics Guides
      • Designing Millifluidics Using Eagle
      • Constructing Laser Cut Millifluidics
      • Constructing Millifluidics via SLA Printing
      • Calibrating IPPs
      • Operating Millifluidics
  • Troubleshooting
    • Troubleshooting Overview
    • Experiment Troubleshooting
      • Contamination
      • Vial Overflow, Pump Failure, and Spills
      • Tubing and Connector Blockage and Bursting
    • Optical Density (OD) Readings
    • Vial Troubleshooting
      • Replacing Photodiodes and LEDs
      • Heating Element
      • Stirring
    • Server Troubleshooting
    • Vial Platform Troubleshooting
      • Motherboard Troubleshooting/Replacement
      • 12V Power Supply Troubleshooting/Replacement
    • GUI Troubleshooting
    • eVOLVER Maintenance
  • Hardware
    • Overview of Hardware Architecture
    • Overview of Fluidics
      • Tubing and connectors
      • Peristaltic Pumps
      • Fluidics box
    • Overview of Millifluidics
      • IPPs (Integrated Peristaltic Pumps)
      • Pressure Regulator
    • Vial Caps
      • Universal Vial Cap Construction Guide (Luer Connectors Only)
      • 5 and 7 Port Nylon Tubing Caps Construction Protocol
    • Smart Sleeve
      • Vial Board
      • 🌪️Stirring
      • Temperature
      • Optical Density
        • OD90 vs OD135
    • Motherboard Layout and Circuitry
      • 🌡️Arduino
      • Sensor/Actuator Board Slots
      • Pulse Width Modulation (PWM) Boards
      • Analog-to-Digital Converter (ADC) Boards
    • Raspberry Pi
    • Chassis
    • Light Blocker / Splash Guard
    • Known Issues
  • Software
    • Overview of Software Architecture
    • DPU
      • Calibration
      • custom_script.py
      • Experiment Data Files
      • eVOLVER.py
    • Arduino
    • Server (Raspberry Pi)
      • Calibration Files
      • Configuration Files (conf.yml)
    • Known Issues
  • Extensions
    • Adding A New Experimental Parameter
      • Power Supply
      • Specific Applications
      • Custom Calibration Code
    • Custom Experiments
      • ePACE
        • ATTiny1634 Writing
        • LUX Board Troubleshooting
      • Morbidostat
      • Extractor Column
        • Extractor Volume Maintenance
        • Experiment Start
          • Sterilizing Extractor Fluidics
          • Setting up your experiment
          • Using the extractor script
        • Extractor Analysis
        • Troubleshooting
        • Example protocols
      • Phototroph Growth
        • Setup Phototroph eVOLVER
        • Photo-eVOLVER Smart Sleeves
          • Photo-eVOLVER Smart Sleeve Construction Guide
        • Experiment Guide
    • Custom Fluidics
      • Adding a Third Pump Rack
      • Bubblers / In-Vial Aeration
        • Bubbler Construction Protocol
        • Bubbler Cleaning Protocol
      • Running the slow pumps
    • min-eVOLVER
      • About
      • min-eVOLVER Construction
        • Parts
        • Construction Protocol
      • Fluidics Setup
      • Software Installation and Startup
      • send_command.py
      • Calibrations
      • Starting an Experiment
      • ePACE with min-eVOLVER
        • [v1.1] ePACE with min-eVOLVER
      • Troubleshooting
    • Interfacing with Other Systems
  • Contributing
    • Guidelines for Contribution
    • Reporting a Bug / Hardware Failure
    • Documentation
      • Making a Forum Post
      • How to Edit the Wiki
    • Software Development
    • Hardware Development
Powered by GitBook
On this page
  • From eVOLVER Forum
  • Overview
  • Diagnosis
  • Replacement Protocol
  • Unfasten Motherboard
  • Remove Motherboard
  • Transfer Parts to New Motherboard
  • Prepare New Motherboard and Plug in to eVOLVER
  • Power Up eVOLVER

Was this helpful?

Edit on GitHub
Export as PDF
  1. Troubleshooting
  2. Vial Platform Troubleshooting

Motherboard Troubleshooting/Replacement

PreviousVial Platform TroubleshootingNext12V Power Supply Troubleshooting/Replacement

Last updated 2 months ago

Was this helpful?

From eVOLVER Forum

Overview

The is the large printed circuit board (PCB) in the vial platform that connects the Smart sleeves to the PWM boards, ADC boards, and SAMD21 mini breakout boards and these boards to the Raspberry Pi. In the event that the motherboard is damaged, it will need to be replaced. Some damage to the motherboard will be immediately obvious.

Diagnosis

to GitHub hardware page for the motherboard

If you have determined that the server and Arduinos are both functioning properly, however, the link between them seems to be broken in some or all ways, then the last likely culprit is the motherboard.

In the example below, overflow events during an experiment caused media to leak through the Smart Vial ribbon cable ports into the vial platform, which houses all of the eVOLVER circuitry. This resulted in visible damage to the motherboard, which can be seen below.

Replacement Protocol

Before getting started with disassembly, be sure to unplug the eVOLVER from the wall outlet.

Unfasten Motherboard

The motherboard is fastened to the eVOLVER platform and other internal components in several ways (numbers correspond to image below):

  1. Ribbon cables connecting the motherboard to the Smart Sleeves (not shown)

  2. Screws in gold holes

  3. 6 screws that fasten wiring from the power supplies to the motherboard

  4. Cable connecting motherboard to Raspberry Pi (fastened to base of platform)

  5. Cable connecting motherboard to display screen (fastened to lid of platform, obstructed by ribbon cable in image)

Remove Motherboard

  1. Unplug the cables connecting the motherboard to the Raspberry Pi and display screen.

  2. Remove the screws in the gold holes using a Phillips screwdriver.

  3. Remove the screws holding the wiring in place using a flathead screwdriver.

  4. Close the eVOLVER chassis lid half way, making sure to hold the motherboard with one hand. Unplug the ribbon cables that connect the Smart Sleeves to the motherboard, starting from the last row of vials and moving towards the first row of vials.

  5. Maneuver the motherboard around the 3D printed part housing the display screen.

Transfer Parts to New Motherboard

The PWM boards, ADC boards, SAMD21 mini Arduino boards and jumpers need to be transferred to the new motherboard.

Be sure to take note of where each board is on the old motherboard before removing them. An image of the configuration can be seen in the images above.

Remove the ADC boards, PWM boards, and SAMD21 mini breakout boards from their slots

  1. Using an Integrated Circuit (IC) puller, as seen in the image below.

    1. Pull the IC puller perpendicular to the motherboard to prevent bending or breaking the SA boards’ pins.

  2. Or put a screwdriver in the hollow under the board and SLOWLY work it back and forth the pry the board out.

Make sure you pull the pins out as straight as possible because it will be challenging to get the board back in if they are bent.

BE VERY CAREFUL. If you break off the wrong connector pin on their bottom, it's pretty annoying to figure out a solution other than replacing the whole board. Unfortunately, you also need to apply a good amount of force to get the board up. But be patient!

  1. If pins are bent in the process of removing the boards, they can be straightened using tweezers. Avoid bending the pins too much; excess bending can cause the pins to snap off.

  2. If pins are snapped off during this process:

    1. Alternatively, the pins can be soldered back into place onto the board, but this is much trickier to do.

    2. The board should be replaced with a new board if other options do not work

Optional: If your motherboard includes a variable resistance board, proceed to steps 3 and 4. If not, skip to step 5.

  1. Remove the variable resistance board from the motherboard. The ADC boards on this board can be removed first, or the variable resistance board and ADC boards can be removed as a single part.

  2. Remove the headers under the variable resistor board and transfer to the new motherboard. These can be removed with pliers.

  1. On the side of the resistor array, there is a description of the part and a black dot, seen in the image below. This black dot indicates the ground pin of the resistor array, and should go into the slot towards the top of the motherboard.

  1. Remove the jumpers to the old motherboard by pulling them straight up. Transfer the jumpers to the new motherboard, recreating the same configuration as shown below.

Before transferring the SA boards to the new motherboard, they should be evaluated for damage. Damage to SA boards will look like burns, melted parts, etc. Damaged SA components should be replaced.

  1. After checking for damage, transfer the variable resistance board (if applicable) and SA boards to the new motherboard (it is easiest to move these boards after transferring the resistor arrays and jumpers). Be sure to align the boards’ pins with the slots on the motherboard before pushing the boards into place.

You may need to apply some force to get the SA boards into their slots. HOWEVER, make sure you are not bending the pins further as you push them in

Adding the SA parts to their slots on the motherboard can be quite tricky, especially if their pins were bent during the removal process. Some tips for this step:

  • Inspect boards and use tweezers to carefully straighten pins that have been bent out of place. Bending the pins back and forth too much can cause them to break off the board, so this should be kept to a minimum.

  • If pins break off an SA board, they can be soldered back into place, or the entire board can be swapped out.

  • Place the SA board over its slots and check for pin alignment. Try to wedge the bent pins into place first, before pressing the board into place.

  • Make sure ALL PINS enter their respective slots before pressing the SA board into place.

  • The SA boards can withstand some force when being pushed into the motherboard. Press the SA boards into their slots until the boards lay flush with the slots on the motherboard.

  1. Unscrew the 3D printed corners from the old motherboard and fasten them to the new motherboard. These parts act as spacers to prevent the backside of the motherboard from touching the eVOLVER platform.

Prepare New Motherboard and Plug in to eVOLVER

  1. Use electrical tape to cover all exposed elements on the back of the motherboard. There are exposed pins on both the top and bottom of the motherboard.

The new motherboard is now ready to be fastened to the vial platform!

  1. Align the motherboard correctly by lining up the ribbon cable ports with the openings in the platform.

  2. Fasten the motherboard to the platform using the screws and a Phillips screwdriver.

  3. Fasten the wiring from the power supplies to the motherboard. The new motherboard will have screws in place for where the wires should go, so those should be removed and then used to fasten the wires to the motherboard.

  4. Plug in the cables to the Raspberry Pi and display screen.

  5. Plug in the Smart Vial ribbon cables to the motherboard from the top of the eVOLVER.

Power Up eVOLVER

You are now ready to power up your eVOLVER!

The first time you turn on the eVOLVER after replacing the motherboard, we recommend opening the platform and keeping an eye out for a few things:

  • Be sure to turn on the 5V power supply first, wait 10 seconds, then turn on the 12V power supply.

  • If you notice smoke or the smell of something burning, immediately turn off both power supplies. If this occurs, inspect the motherboard and the SA boards for melting or burn spots.

  • The red light on the Controls Raspberry Pi should turn on steady, and the green light should blink on and off.

  • The red and blue lights on the SAMD21 mini breakout boards should turn on steady.

You may want to purchase an to help remove the PCBs from the motherboard without damaging them.

Not all pins are needed. You can check the hardware KiCad for whether the pin you snapped is necessary

Remove the yellow resistor arrays (see below). An or pliers may be helpful here. Transfer the resistor arrays to the new motherboard, making sure to insert them in the proper orientation.

Integrated Circuit (IC) puller
files
IC puller
12V Power Supply Connection to Motherboard Melted
motherboard
Link
Layout of damaged motherboard
Internal components fastening the motherboard to the vial platform
Using an IC puller to remove an ADC board. In black are the PWM boards. Small red boards are the SAMD21 Arduinos
Variable resistance board sitting on top of headers after being removed
Jumper configuration indicating power supply for each component
3D printed spacers in corners